Aerodynamic Particle Sizer
Features
1.Sampling flow rate of 0.5L/min is ideal for particle size analysis in animal exposure tests.
2.Sampler is divided into 7 stages + membrane stage.
Application
1.Aerodynamic particle size tests in aerosol inhalation toxicology studies.
2.Aerodynamic particle size tests of particles in environment.
3.Aerodynamic particle size tests of dust in occupational health workplaces.
Specification
| Stage | Cutting Diameter (μm) |
| 1 | 5.1 |
| 2 | 3.1 |
| 3 | 1.9 |
| 4 | 1.2 |
| 5 | 0.77 |
| 6 | 0.50 |
| 7 | 0.33 |
Specification
|
Sampling principle |
Inertial impaction |
|
Sampling flow rate |
0.5L/min |
|
Collection plate material |
SUS 304 |
|
Collection plate size |
22 mm diameter |
|
Number of stages |
7 stages + filter stage |
|
Instrument dimensions |
φ44.5mm x 164mm |
|
Instrument weight |
0.8kg (accessories excluded) |