Aerosol Mass Concentration Detector

Model HRH-AMC01 Aerosol mass concentration detector is mainly used to measure mass concentration of aerosols, dust, smoke and other substances. Different from traditional weighing methods, the detector does not need a long sampling time and steps such as balance weighing, which greatly simplifies operation, and can be used for real-time evaluation of mass concentration trends. Compact and portable design enables selections of handy sampling and bracket fixed sampling. Based on principle of near forward light scattering, scattered light generated when aerosol passes through detection chamber is captured by optical detector and converted into photoelectric signals, and then mass concentration is displayed on screen in real time.

    Features

    1. Measuring range: up to 250g/m3;

    2. Humanized structural design, easy to carry, operate and measure;

    3. Unique calibration plug design for site calibration;

    4. Set the correction factor to perform factor correction for specific substances;

    5. "Data curve" mode, observe data change trends in time;

    6. Two power supply methods: alkaline batteries and USB DC5V.

    Application

    1.Monitoring of aerosol concentration. 

    2.Risk assessment of dust and suspended matter.

    3.Monitoring of dust concentrations in workplace. 

    4.Industrial process monitoring.

    5.Environmental dust assessment. 

    6.Boundary monitoring in construction industry.

    Working Principle

    In clean gases, all light emitted by laser source will be blocked by light trap and cannot reach optical detector. When aerosol particles enter measuring chamber, light beam irradiated on particle surface or aerosol substances will produce near forward scattering. After action of lens group, part of scattered lights can be transmitted to optical detector for photoelectric conversion. The detector adopts near forward scattering principle, it will not be too sensitive to particles refractive index and material color itself, which improves detection accuracy.

    Aerosol Mass Concentration Detector

    As shown in figure, aerosol mass concentration detector mainly consists of signal emission, reception, processing, LCD display, keyboard operation, and data transmission, etc. Firstly, laser source is turned on by signal emission unit, then there are signals received, signal processing unit processes received signal and converts into aerosol mass concentration according to parameters, finally mass concentration value will be transmitted and displayed.

    Specification

    Application scope 1μg/m³ - 250g/m³
    Resolution 1μg/m³
    Photosensitive technique Forward light scattering - 650nm
    Display screen 320*240 TFT colored LCD screen
    Keyboard Eight-key keyboard
    Ambient temp 0°C- 50°C (non-condensing)
    Power supply AA 1.5V*3 alkaline batteries; DC-5V type-c USB
    Communication RS485 Modbus protocol, supporting network communication for up to 99 detectors

    Standards

    JJG 846-2015 Dust Concentration Measuring Instruments