Biosafety Nose-only Exposure System for Non-human Primates - China Suppliers and Factory for Inhalation Testing
Features
Perform inhalation exposure tests for 3 monkeys simultaneously.
Class III BSC cabinet: compact design in stainless steel, excellent corrosion resistance, large round corners design is easy to clean. Double-sided design allows multi-users to operate together.
Integrated safety design: safety program coordinated control, sterilization and purification program settings, corrosion-resistant sensors, double HEPA filtration, intelligent pressure difference control, internal and external interlock, startup self-test, etc.
Negative pressure inside cabinet is not less than 120Pa, and mask pressure is not less than 100Pa.
Exhaust HEPA filter is Class H14 with leakage detection and sterilization in place.
HMI has functions of parameter settings, real-time display and data recording. Real-time monitor parameters of oxygen concentration, carbon dioxide concentration, temp, humidity, pressure (between cabinet and HEPA filter), and abnormal alarms.
One-to-one control of monkey mouth and nose inhalation, enhances dose stability, accuracy and reproducibility.
Translucent aerosol generator enables visual monitoring of liquid level and atomization. Liquid samples can be added midway without downtime.
Monkey chair fits 3-6kg monkeys with adjustable sitting angle which makes monkey feel better.
The seat and back of the monkey chair are made of PTFE for comfort. A removable waste collector is located underneath the seat panel.
Monkeys can be quickly bound by velcro tape, which is convenient and reliable.
Application
Applicable to inhalation exposure tests of inhalation poisoning, inhalation immunity, R&D of inhalation formulations in ABSL-3, ABSL-4 laboratories.
Working Principle
The Biosafety Nose-only Exposure System for Non-human Primates is nose-only aerosol inhalation infection equipment for rodents combined with high-level biosafety protection. Key components are nose-only exposure system and Class III BSC cabinet. Applicable to nose-only exposure tests for rodents (e.g., mice, rats, guinea pigs, ferrets, rabbits, etc.) with same dose under non-anesthesia conditions. Meanwhile, Class III BSC cabinet keeps personnel, laboratory and environment safe during virulent pathogens tests.
Airflow Direction
Specification
Cabinet Dimensions: < 3100*900*2250mm (Working area: 1794*835*900mm)
Monkey Chair Angle: Can be adjusted 90°
Extraction & Sampling Flow: 1-30L/min (adjustable)
Monkey Chair Size: 340*300*600 mm
Cabinet Negative Pressure: ≥ 120Pa
Face Mask Pressure: ≥ 100Pa
Aerosol Generation Flow Rate: > 0.2 mL/min
Capacity: Inhalation test for 3 monkeys
Microbial Aerosol Error: Stability & evenness error < 20%
Standards
RB∕T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment
OECD TG 403 Acute Inhalation Toxicity (2009)
OECD TG 436 Acute Inhalation Toxicity. Acute Toxic Class Method (2009)
OECD TG 412 28d Sub-acute inhalation (2018)
OECD TG 413 90d Sub-chronic inhalation (2018)
Frequently Asked Questions
The system is specifically designed to perform inhalation exposure tests for 3 monkeys simultaneously.
This system is applicable to inhalation poisoning tests, inhalation immunity tests, and R&D of inhalation formulations in ABSL-3 and ABSL-4 laboratories.
The negative pressure inside the Class III BSC cabinet is maintained at not less than 120Pa, and the face mask pressure is maintained at not less than 100Pa.
The system complies with RB∕T 199-2015, and multiple OECD guidelines including OECD TG 403 (Acute Inhalation Toxicity), OECD TG 436 (Acute Toxic Class Method), OECD TG 412 (28d Sub-acute inhalation), and OECD TG 413 (90d Sub-chronic inhalation).
The HMI monitors real-time parameters of oxygen concentration, carbon dioxide concentration, temperature, humidity, pressure (between the cabinet and HEPA filter), and provides abnormal alarms.