Biosafety Nose-only Exposure System for Non-human Primates - China Suppliers & Factory for Inhalation Tests
Features
- Perform inhalation exposure tests for 3 monkeys simultaneously.
- Class III BSC cabinet: compact design in stainless steel, excellent corrosion resistance, large round corners design is easy to clean. Double-sided design allows multi-users to operate together.
- Integrated safety design: integrated design of Class III BSC cabinet and exposure tower mainly includes safety program coordinated control, sterilization and purification program settings, corrosion-resistant sensors, double HEPA filtration, intelligent pressure difference control, internal and external interlock, startup self-test, etc.
- Negative pressure inside cabinet is not less than 120Pa, and mask pressure is not less than 100Pa.
- Exhaust HEPA filter is Class H14 with leakage detection and sterilization in place.
- HMI has functions of parameter settings, real-time display and data recording, etc. Real-time monitor parameters of oxygen concentration, carbon dioxide concentration, temp, humidity, pressure (between cabinet and HEPA filter), and abnormal alarms.
- One-to-one control of monkey mouth and nose inhalation, enhances dose stability, accuracy and reproducibility.
- Translucent aerosol generator enables visual monitoring of liquid level and atomization. Liquid samples can be added midway without downtime.
- Monkey chair fits 3-6kg monkeys with adjustable sitting angle which makes monkey feel better.
- The seat and back of the monkey chair are made of PTFE for comfort. A removable waste collector is located underneath the seat panel.
- Monkeys can be quickly bound by velcro tape, which is convenient and reliable.
Application
Applicable to inhalation exposure tests of inhalation poisoning, inhalation immunity, R&D of inhalation formulations in ABSL-3, ABSL-4 laboratories.
Working Principle
The Biosafety Nose-only Exposure System for Non-human Primates is nose-only aerosol inhalation infection equipment for rodents combined with high-level biosafety protection. Key components are nose-only exposure system and Class III BSC cabinet. Applicable to nose-only exposure tests for rodents (e.g., mice, rats, guinea pigs, ferrets, rabbits, etc.) with same dose under non-anesthesia conditions. Meanwhile, Class III BSC cabinet keeps personnel, laboratory and environment safe during virulent pathogens tests.
Specification
Standards
- RB∕T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment
- OECD TG 403 Acute Inhalation Toxicity (2009)
- OECD TG 436 Acute Inhalation Toxicity. Acute Toxic Class Method (2009)
- OECD TG 412 28d Sub-acute inhalation (2018)
- OECD TG 413 90d Sub-chronic inhalation (2018)
Frequently Asked Questions (FAQ)
The system is designed to perform nose-only inhalation exposure tests for up to 3 monkeys simultaneously with one-to-one control to ensure dose stability and accuracy.
The Class III BSC cabinet operates under negative pressure (not less than 120Pa) and features double H14 HEPA exhaust filtration with in-place sterilization and leakage detection to protect personnel and the laboratory environment.
Yes, the translucent aerosol generator allows you to visually monitor liquid levels and add liquid samples midway through the process without needing to shut down the system.
The comfortable PTFE-based monkey chair is designed to fit monkeys weighing between 3kg and 6kg, featuring an adjustable sitting angle and a removable waste collector.
The system complies with multiple international standards, including RB∕T 199-2015, OECD TG 403 (Acute Inhalation Toxicity), OECD TG 436, OECD TG 412 (28d Sub-acute), and OECD TG 413 (90d Sub-chronic).