Biosafety Nose-Only Exposure System for Rodents by China Suppliers | High-Safety Factory Equipment
Features
- Animal adaptability: mice, rats, guinea pigs, ferrets, rabbits etc.
- Exposure environment monitoring: exposure room temp, humidity, pressure difference, carbon dioxide concentration, and oxygen concentration can be monitored in real-time during tests, compliant with the requirements of inhalation guidelines.
- Parameters real-time monitoring: aerosol generation flow rate, dilution flow rate, sampling flow rate, pumping flow rate, poisoning time, remaining time, etc.
- Adopt a high-precision gas mass flow controller, which fully meets the requirements of gas flow accuracy in inhalation exposure tests.
- Class III BSC cabinet: compact design in stainless steel, excellent corrosion resistance, large rounded corners ensure no cleaning of dead corners. Double-sided design allows multi-users to operate together.
- Integrated safety design: integrated design of the Class III BSC cabinet and exposure tower mainly includes safety program coordinated control, sterilization and purification program setting, corrosion-resistant sensors, double HEPA filtration, intelligent pressure difference control, internal and external interlock, startup self-test, etc.
- Advanced technology platform: simulate natural infection of respiratory pathogen, provide an advanced evaluation platform for pathogenic microorganism inhalation infection, vaccine aerosol immunity, lung drug delivery and aerosol challenge evaluation for vaccine or drug.
- Nose-only inhalation infection: avoid other exposure pathways such as digestive tract and skin, which conform to the natural infection pathway of respiratory pathogens.
- Concentration uniformity and stability: adopt orthogonal dilution patented technology (CN201310351052.9), and consider hydrodynamics and other key parameters comprehensively to ensure microbial concentration is stable and uniformly distributed. The stability and uniformity of microbial aerosol deviation is less than ±20%.
- Software function: equipped with functions of user classified management, audit trail, data unerasable, etc., which meet requirements of GLP laboratories. It can carry out parameter settings, real-time monitoring and environmental parameters recording during tests, as well as real-time alarms and recording of abnormalities.
- Test data output: all experimental data can be stored and accessed, and reports can be issued automatically after test.
Application
Exposure tests of inhalation infection, inhalation immunity, inhaler R&D in ABSL-3 and ABSL-4 laboratories.
Working Principle
The Biosafety Nose-only Exposure System for Rodents mainly consists of bioaerosol generator, monkey chair, dedicated negative pressure cabinet for aerosol infections, multifunctional exposure mask, exposure environmental detection and control system, host controller, gas controller, bioaerosol sampler, aerosol mass concentration detector, bioaerosol waste gas treatment system, and protection alarm system. The negative pressure environment is kept through Class III BSC cabinet, which is isolated from laboratory by means of inlet HEPA filter and double HEPA filtration, airflow direction is shown in the figure.
Airflow Direction
Exposure Tower for Rodents
Specification
| Parameter | Specification Details |
|---|---|
| Dimensions of Class III BSC cabinet | 2400*900*2100mm |
| Dimensions of Class III BSC cabinet working area | 1794*835*900mm |
| Operation sides of Class III BSC cabinet | Double-sided |
| Filter precision | H14/99.99% |
| Noise | Lower than 60db |
| Exposure ports of 2 layers exposure tower | 24 ports |
| Flow rate of 3-hole aerosol generator | 6L/min |
| Particle size of liquid aerosol | MMAD1-3m, GSD:1.5-3 |
| Humidity of exposure chamber during test | 30-70%RH |
| Oxygen concentration in exposure chamber | 19-21% |
| Carbon dioxide concentration in exposure chamber | Less than 1% |
Liquid Aerosol Generator
| Model | Nozzle Number | Rated Flow L/min | Normal Operating Flow Range L/min |
|---|---|---|---|
| HRH-WAG1 | 1 | 2 | 1-2 |
| HRH-WAG3 | 3 | 6 | 3-6 |
| HRH-WAG6 | 6 | 12 | 6-12 |
| HRH-WAG12 | 12 | 24 | 12-24 |
| HRH-WAG24 | 24 | 48 | 24-48 |
Standards
- RB∕T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment
- OECD TG 403 Acute Inhalation Toxicity (2009)
- OECD TG 436 Acute Inhalation Toxicity. Acute Toxic Class Method (2009)
- OECD TG 412 28d Sub-acute inhalation (2018)
- OECD TG 413 90d Sub-chronic inhalation (2018)