Breathing Simulator HRH-BRM2100 for Nebulizer Testing - Top China Suppliers & Factory Compliance with USP, Ph.Eur
Features
USP, Ph. Eur. and ChP 2020 compliant.
Tidal volume setting range is 0-1000ml.
Built-in parameters for four standard respiratory modes specified in Pharmacopoeia including standard adult, standard child, standard infant and standard neonate.
User-defined parameters: tidal volume, respiratory rate, I/E ratio, respiratory pattern curve, cycle number, etc. could be set to meet different test needs.
English/Chinese switchable interface is convenient for operation.
The parameters are calculated automatedly, also curves of breath flow volume and tidal volume can be displayed in real-time.
Application
Specification
| Dimensions | 70x35x62cm |
| Weight | 35Kg |
| Power rate | 0.85KW |
| Max. Tidal Volume | 1000ml |
| Precision | Max. TV±5% or Max. TV±1ml |
| Power supply | 220V, 50Hz |
Different Breath Characteristics of Breath Simulator
| Item | Adult | Neonate | Infant | Child |
|---|---|---|---|---|
| Tidal Volume | 500ml | 25ml | 50ml | 155ml |
| Respiratory Rate | 15 cycles/min | 40 cycles/min | 30 cycles/min | 25 cycles/min |
| Breathing Profile Curve | Sine type | Sine type | Sine type | Sine type |
| I/E ratio | 1:1 | 1:3 | 1:3 | 1:2 |
Environmental Requirements
Working Environment
Temperature: 0°C-60°C
Humidity: 5% -95% (non-condensing state)
Storage Environment
Temperature: -20°C-70°C
Humidity: 5% -95% (non-condensing state)
Components
- Drive module: High-precision servo motor with full closed-loop control, 4M max input frequency, friction compensation, and anti-collision function.
- Transmission module: High-precision ball screw offering low friction, high efficiency and accuracy, high axial stiffness, and low backlash.
- Motion module: High-precision stainless steel piston and cylinder, powered by an AC servo unit with DSP, FPGA, and PIM for vibration suppression.
- Control module: Laser galvanometer motion controller (DSP + FPGA) with a 250μs control cycle and up to 1000 data points per second.
- Display module: Real-time display of flow rate and tidal volume curves.
Compliance
Frequently Asked Questions (FAQ)
What is the tidal volume setting range for this breath simulator?
The tidal volume setting range of the simulator is 0 to 1000 ml, allowing it to adapt to various experimental requirements.
Which pharmacopoeia standards does the device comply with?
The device is fully compliant with USP, Ph. Eur., and ChP 2020 standards, ensuring reliable regulatory compliance for Nebulizer DDU tests.
What standard respiratory modes are built into the system?
It includes pre-loaded parameters for four standard respiratory modes specified in pharmacopoeias: standard adult, standard child, standard infant, and standard neonate.
Can users customize their own breathing parameters?
Yes, user-defined parameters such as tidal volume, respiratory rate, I/E ratio, respiratory pattern curve, and cycle number can be set to meet specific test needs.
What are the operating and storage environmental requirements?
The working environment requires a temperature range of 0°C to 60°C and humidity of 5% to 95% (non-condensing). The storage environment supports temperatures from -20°C to 70°C and humidity of 5% to 95% (non-condensing).
How does the host system ensure motion precision?
The system utilizes a high-precision servo motor with full closed-loop control, a zero-backlash ball screw, and a laser galvanometer motion controller with a 250μs control cycle to eliminate volumetric lag and ensure high fidelity.