Request a Quote
Leave Your Message

Breathing Simulator

Model HRH-BRM2100 breathing simulator is designed for testing nebulizers. It can meet instrument qualifications of USP and Ph.Eur., as well as determination of delivery rate and delivered dose of nebulizers in ChP 2020.

    Features

    USP, Ph. Eur. and ChP 2020 compliant.

    Tidal volume setting range is 0-1000ml.

    Built-in parameters for four standard respiratory modes specified in Pharmacopoeia including standard adult, standard child, standard infant and standard neonate.

    User-defined parameters: tidal volume, respiratory rate, I/E ratio, respiratory pattern curve, cycle number, etc. could be set to meet different test needs.

    English/Chinese switchable interface is convenient for operation.

    The parameters are calculated automatedly, also curves of breath flow volume and tidal volume can be displayed in real-time.

    Application

    For Nebulizer DDU Tests; Automated Simulated Extraction of Total Emitted Dose from Inhaler Devices.

    In the in vitro evaluation of inhalation drug products, accurately replicating the patient's real‑world breathing pattern is the key to assessing the device's true drug delivery performance. The HRH‑BRM2100 incorporates an exceptionally high‑order precision drive architecture. The system is equipped with an ultra‑low friction 1‑L precision cylinder, driven by an industrial‑grade high‑torque servo motor coupled with a zero‑backlash ball‑screw slide table. This rigid mechanical linkage translates the motor's millisecond‑scale rotational motion into the piston's absolute linear displacement with 100% fidelity, completely eliminating volumetric lag.

    Designed for standard routine testing scenarios, this 1L breath simulator features a powerful micro‑controller system pre‑loaded with all standard physiological breathing parameters as specified in pharmacopoeias. Users can operate the device via a keyboard and mouse, inputting tidal volume (0–1000 mL), respiratory rate, and inspiratory‑to‑expiratory (I/E) ratio according to experimental requirements. In addition, the system's built‑in high‑speed fluid algorithm dynamically plots real‑time flow and tidal volume waveforms, providing full visibility into the airflow dynamics and equipment status throughout the entire test cycle.

    Specification

    Dimensions 70x35x62cm
    Weight 35Kg
    Power rate 0.85KW
    Max. Tidal Volume 1000ml
    Precision Max. TV±5% or Max. TV±1ml
    Power supply 220V, 50Hz


    Different Breath Characteristics of Breath Simulator

    Item

    Adult

    Neonate

    Infant

    Child

    Tidal Volume

    500ml

    25ml

    50ml

    155ml

    Respiratory Rate

    15 cycles/min

    40 cycles/min

    30 cycles/min

    25 cycles/min

    Breathing Profile Curve

    Sine type

    Sine type

    Sine type

    Sine type

    I/E ratio

    1:1

    1:3

    1:3

    1:2

    Working environment

    Temperature: 0°C-60°C
    Humidity: 5% -95% (non-condensing state) 

    Storage environment

    Temperature: -20°C-70°C
    Humidity: 5% -95% (non-condensing state)

    Components

    The instrument consists of a filter sampler and a breath simulator host. The filter sampler includes a filter paper and holder. The low-resistance filter paper enables quantitative aerosol collection and active substance recovery using appropriate solvents. The filter holder has a dead volume of less than 10% of the standard adult tidal volume.
    The breath simulator host comprises five modules:
    1.Drive module: High-precision servo motor with full closed-loop control, 4M max input frequency, friction compensation, and anti-collision function.
    2.Transmission module: High-precision ball screw offering low friction, high efficiency and accuracy, high axial stiffness, and low backlash.
    3.Motion module: High-precision stainless steel piston and cylinder, powered by an AC servo unit with DSP, FPGA, and PIM for vibration suppression.
    4.Control module: Laser galvanometer motion controller (DSP + FPGA) with a 250μs control cycle and up to 1000 data points per second.
    5.Display module: Real-time display of flow rate and tidal volume curves.Components

    Compliance

    USP, Ph.Eur. and ChP 2020

    光影魔术手拼图2

    Typical Users

    pure pharmaceutical manufacturing L.L.C
    Shenzhen Wuxin Technology Co., Ltd
    Guangdong Medical Instrument Institute
    Vkan Certification & Testing Co., Ltd.
    Henan PURUI Pharmaceutical Co., Ltd.
    CCIC Huatongwei International Inspection (Suzhou) Co., Ltd.
    Shanghai Fosun Pharmaceutical Industry Development
    Chengdu Institute Food and Drug Control
    Suzhou Inhal Pharma Co., Ltd.
    Beijing Weishi Rui Medical Tech Co Ltd
    Zhengzhou Autu Industrial Group Co., Ltd.
    Jiangsu Chia Tai Fenghai Pharmaceutical Co., Ltd.
    Jiangsu Chiatai Qingjiang Pharmaceutical Co., ltd
    Nanchang University
    Shanghai Saizeng Medical Technology Co., Ltd.
    TÜV SÜD Certification and Testing (China) Co., Ltd.
    Shenzhen Aoji Health Technology Co., Ltd.
    Guangzhou Grg Metrology & Test Co., Ltd.
    China State Institute of Pharmaceutical Industry (Haimen) R&D Co., Ltd.
    Shenzhen Jermei Medical Device Technology Co., Ltd.
    Nanjing JSIAMA Biopharmaceutical Co., Ltd.
    China State Institute of Pharmaceutical Industry (Haimen) R&D Co., Ltd.
    China Coal Research Institute (CCRI)

    Make An AppointmentLEAVE YOUR MESSAGE