China HRH-BRM2100 Breathing Simulator for Testing Nebulizers - Leading Suppliers and Factory Solutions
Features
USP, Ph. Eur. and ChP 2020 compliant.
Tidal volume setting range is 0-1000ml.
Built-in parameters for four standard respiratory modes specified in Pharmacopoeia including standard adult, standard child, standard infant and standard neonate.
User-defined parameters: tidal volume, respiratory rate, I/E ratio, respiratory pattern curve, cycle number, etc. could be set to meet different test needs.
English/Chinese switchable interface is convenient for operation.
The parameters are calculated automatedly, also curves of breath flow volume and tidal volume can be displayed in real-time.
Application
For Nebulizer DDU Tests
In the in vitro evaluation of inhalation drug products, accurately replicating the patient's real‑world breathing pattern is the key to assessing the device's true drug delivery performance. The HRH‑BRM2100 incorporates an exceptionally high‑order precision drive architecture. The system is equipped with an ultra‑low friction 1‑L precision cylinder, driven by an industrial‑grade high‑torque servo motor coupled with a zero‑backlash ball‑screw slide table. This rigid mechanical linkage translates the motor's millisecond‑scale rotational motion into the piston's absolute linear displacement with 100% fidelity, completely eliminating volumetric lag.
Designed for standard routine testing scenarios, this 1‑L breath simulator features a powerful micro‑controller system pre‑loaded with all standard physiological breathing parameters as specified in pharmacopoeias. Users can operate the device via a keyboard and mouse, inputting tidal volume (0–1000 mL), respiratory rate, and inspiratory‑to‑expiratory (I/E) ratio according to experimental requirements. In addition, the system's built‑in high‑speed fluid algorithm dynamically plots real‑time flow and tidal volume waveforms, providing full visibility into the airflow dynamics and equipment status throughout the entire test cycle.
Specification
| Dimensions | 70x35x62cm |
| Weight | 35Kg |
| Power rate | 0.85KW |
| Max. Tidal Volume | 1000ml |
| Precision | Max. TV±5% or Max. TV±1ml |
| Power supply | 220V, 50Hz |
| Item | Adult | Neonate | Infant | Child |
|---|---|---|---|---|
| Tidal Volume | 500ml | 25ml | 50ml | 155ml |
| Respiratory Rate | 15 cycles/min | 40 cycles/min | 30 cycles/min | 25 cycles/min |
| Breathing Profile Curve | Sine type | Sine type | Sine type | Sine type |
| I/E ratio | 1:1 | 1:3 | 1:3 | 1:2 |
Working environment
Temperature: 0°C-60°C
Humidity: 5% -95% (non-condensing state)
Storage environment
Temperature: -20°C-70°C
Humidity: 5% -95% (non-condensing state)
Components
The instrument consists of a filter sampler and a breath simulator host. The filter sampler includes a filter paper and holder. The low-resistance filter paper enables quantitative aerosol collection and active substance recovery using appropriate solvents. The filter holder has a dead volume of less than 10% of the standard adult tidal volume.
The breath simulator host comprises five modules:
Compliance
USP, Ph.Eur. and ChP 2020
Frequently Asked Questions (FAQ)
Which pharmacopoeia standards does the breath simulator comply with?
The system is fully compliant with USP, Ph. Eur., and ChP 2020 regulatory requirements for inhalation drug testing.
What is the range of the tidal volume settings?
The tidal volume setting range is from 0 to 1000 ml, allowing for a wide range of test configurations.
Are there preset respiratory modes for different patient groups?
Yes, it features built-in parameters for four standard respiratory modes specified in the Pharmacopoeia: standard adult, standard child, standard infant, and standard neonate.
Can I customize my own breathing cycle parameters?
Yes, users can set custom parameters including tidal volume, respiratory rate, I/E ratio, respiratory pattern curve, and cycle number to meet specific testing needs.
Does the system support real-time data monitoring?
Yes, the system automatically calculates parameters and displays real-time curves for both breath flow volume and tidal volume during operation.