China HRH-VOC186 VOC Generator - Reliable Suppliers & Factory for Controlled Concentration of Volatile Organic Compounds

Introducing the HRH-VOC186 VOC generator, a state-of-the-art instrument designed for the precise and rapid evaporation of volatile organic compounds (VOCs), such as alcohol and formaldehyde. This innovative device allows for the generation of gas at controlled concentrations, making it an essential tool for various applications in environmental monitoring and industrial processes. As a reliable supplier based in China, our factory ensures top-notch quality and performance in every unit, catering to the needs of customers seeking efficient VOC solutions

    Features

    • Gas concentration generated is controllable and with high continuous stability.
    • Adopt glass ball stack convective volatilization structure with fast volatilization speed.
    • Assist heating volatile gas, temp can be controlled.
    • Generating system is closed with high safety.
    • Volatilization rate: 0-5mL/min, gas flow rate: 0-20L/min, heating temp: 20-60℃.

    Application

    • Inhalation Toxicity Tests on VOC.
    • Health Risk Assessment of Environmental Atmosphere and Indoor Volatile Solvents.
    • Inhalation Toxicity Studies of VOC or Petrochemicals in Occupational Sites.
    • Online Monitor and Control for VOC.

    Specification

    Volatilization rate 0-5mL/min
    Gas flow rate 0-20L/min
    Heating temperature 20-60℃

    Working principle

    The volatile sample is quantitatively injected into the volatilization pipeline filled with glass beads through the sample transfer pump, the superficial area of the dropped sample is increased enough to allow for faster volatilization and the clean air is heated by the air heater to enter the cushion chamber, and then the sample volatilization is promoted by the glass sphere pile and volatilized samples and air can be imported into the exposure system through the outlet for contamination. There is a temperature probe at the outlet of VOC gas for detecting the temperature of the outlet gas, the temperature value is fed back to the air heater, and the heater is used to intelligently adjust the heating power according to the outlet temperature set by the user.
    Working principle
    Note:
    1. VOC concentration = liquid sample mass/ (liquid sample gasification expansion volume + air volume);
    When the experimental concentration of VOC is designed to be relatively low, the concentration of VOC is ≈ liquid sample mass/air volume;
    2. The temperature value set by the air heater is the temperature of the air outlet of the VOC system other than the hot air temperature after heating;
    3. To be able to quantify the VOC content, the injection rate of the sample cannot exceed the maximum volatilization rate, that's to say that it's necessary to ensure that the sample injected into the volatilization tube is completely volatilized, and no liquid flows into the cushion chamber.
    4. The temperature setting method is as follows: The temperature control mode is that the system will automatically stop the air heating when the air outlet temperature is greater than the set upper limit; the system will automatically enable the air heating when the temperature is less than the set lower limit.

    Frequently Asked Questions

    What is the volatilization rate and gas flow rate of this system?

    The system features a volatilization rate of 0-5mL/min and a gas flow rate of 0-20L/min.

    How is the heating temperature controlled?

    The heating temperature ranges from 20-60℃. It is controlled intelligently by an air heater that adjusts power based on feedback from a temperature probe located at the outlet of the VOC gas.

    What is the formula for calculating VOC concentration?

    VOC concentration is calculated as: liquid sample mass / (liquid sample gasification expansion volume + air volume). When the designed concentration is relatively low, it can be simplified to approximately: liquid sample mass / air volume.

    How does the automatic temperature control mode work?

    The system automatically stops the air heating when the air outlet temperature exceeds the set upper limit, and automatically enables the heating when the temperature drops below the set lower limit.

    What precautions should be taken to ensure complete sample volatilization?

    To accurately quantify the VOC content, the injection rate of the sample must not exceed the maximum volatilization rate. This ensures the sample in the volatilization tube is completely volatilized and prevents liquid from flowing into the cushion chamber.

    What are the primary applications of this VOC generation system?

    It is mainly used for inhalation toxicity tests on VOC, health risk assessments of environmental and indoor volatile solvents, inhalation toxicity studies in occupational sites, and online monitoring and control for VOC.