Fine Particle Dose Measurement for Nasal Sprays - China Suppliers & Factory Solutions

,

For optimal drug deposition in the nasal cavity instead of the lungs, nasal products like nasal sprays are engineered to produce particles with aerodynamic diameters primarily between 20-200 µm. However, it is crucial to note that around 5% (or less) of these particles may still fall below 10 µm and be deposited in the lower respiratory tract and lungs. Therefore, measuring the fine particle dose (FPD) generated by nasal sprays is essential. In China, suppliers leverage cascade impactors to effectively capture 0-10 µm particles, ensuring accurate assessments. By combining a cascade impactor with a nasal spray expansion chamber, our factory is able to provide precise measurements of the fine particle doses for drugs in nasal sprays and nasal aerosols, enhancing product efficacy and safety.

,

    ⚙️Features

    Cascade impactors are used to capture 0-10μm particles. The combination of cascade impactor with nasal spray expansion chamber is used to measure fine particle dose of drugs in nasal sprays and nasal aerosols.

    🎯Application

    Nasal Spray FPD Tests

    📋Specification

    • 1L capacity: used in nasal aerosols to maximize deposition below the top stage.
    • 2L capacity: used in nasal sprays to maximize nebulization and deposition.
    • 5L capacity: for some stronger nasal sprays, additional room is required to be able to generate aerosol plume.
    Notes: not mandatory to use the expansion chamber for Nasal Powder FPD testing, as its test method is the same as the Dry Powder Inhalers by equipping the same test instrument and accessories, no pre-separator is required.

    🔧Accessories

    Cascade Impactor Accessories

    💻Inhaler Testing Data Analysis Software Instruction

    The main interface of the analysis software is shown in Figure 20. After inputting flow, the software will automatically calculate the cutting particle size. The information related to the experiment can be filled in on the right. The current experiment information can be saved by the Save button. Open the previously saved experiment information by the Open button and click the Next button to enter the data input interface after confirming the flow rate.

    In the data input interface, each column represents an experiment, and each row corresponds to a different stage of the device. The number of experiments can be selected on the right side of the data sheet, and open data files and save data as files by Open and Save Button. After data input is completed, click the Next button to enter the result display interface. The result display interface can display parameters (such as MMAD and GSD) of each experiment as well as particle size distribution fitting curve and scatter diagram. It's allowed to choose to display the curve of any experiment or the curve of all experiments. Click the right mouse button in terms of graphs to export the curve graph.

    Inhaler Testing Data Analysis Software Main Interface
    Inhaler Testing Data Analysis Software Result Interface

    🛡️Compliance

    • FDA Bioavailability and Bioequivalence Studies for Nasal Aerosols and Nasal Sprays for Local Action (2003)
    • FDA Nasal Spray, Inhalation Solution, Suspension & Spray Drug Products(2002)
    • Nasal Preparations (Dosage Forms 0676) Ph.Eur. 2021 10.5

    👥Typical Users

    United Arab Emirates University
    Genix Pharma Private Limited
    Drug International Limited
    Endocrine Technologies Llc
    Novel Phase
    Divine Healthcare Pvt. Ltd.
    Cipla Ltd.
    JOINN Laboratories
    Pharmaron Beijing Co., Ltd.
    Boji Medical Technology Co., Ltd.
    Tianjin Pharmaceutical Research Institute Co., Ltd.
    China State Institute of Pharmaceutical Industry (Haimen) R&D Co., Ltd.
    Shenyang SYRICI Testing Co., Ltd (SYRICIT)
    Guangdong Zhongke Yinghai Tech Co., Ltd.
    Jiangsu Medicine Research Institute
    Chinese Academy of Sciences Shanghai Institute of Materia Medica
    Shenyang Pharmaceutical University Wuya College of Innovation
    Jiangxi University of Chinese Medicine
    Kunming Institute of Zoology. Cas
    Zhongshan Research Institute, Shanghai Institute of Materia Medica
    ZhongShan Institute for Drug Discovery
    Shanghai Xiandai Medicine Preparation Engineering Research Centre Co.,Ltd.
    Northeastern University
    China Institute for Radiation Protection
    Hubei Provincial Center for Disease Control and Prevention
    Inner Mongolia Medical University
    Weifang Medical University
    Suzhou Jingzida Testing Service Co., Ltd.
    Suzhou Inhal Pharma Co., Ltd.
    Zhejiang Cdmo Pharmaceutical Co., Ltd
    Hangzhou Zhezhong Pharmaceutical Technology Co., Ltd.
    Jinan Kangqiao Medical Technology Co., Ltd.
    Anhui JieXi Pharmaceutical Co., Ltd.
    Beijing Xingyuanxiang Biotechnology (Group) Co., Ltd.
    Hunan Prima Pharmaceutical Research Center Co., Ltd.
    Beijing Wehand-bio Pharmaceutical Co.,ltd
    Jinan Baohede Pharmaceutical Technology Co., Ltd.
    Shandong Innovative Drug Research and Development Co., Ltd.
    Zhongshan Liansi Chemical Technology Co., Ltd.
    Jiangxi Zhonglu Science & Technology Co., Ltd.
    Shanghai Zezheng Biopharmaceutical Co., Ltd.
    Guangzhou Puzhiyun Information Technology Co., Ltd.
    Guangzhou Coslan Scientific Instrument Co., Ltd.
    Hainan Dayi Scientific Instrument Co., Ltd.

    ❓Frequently Asked Questions (FAQ)

    What particle size range can the cascade impactor capture?

    Cascade impactors are specifically designed to capture fine particles within the range of 0 to 10μm.

    Why combine a cascade impactor with a nasal spray expansion chamber?

    This combination is essential for measuring the fine particle dose (FPD) of drugs delivered via nasal sprays and nasal aerosols.

    How do I choose between the 1L, 2L, and 5L expansion chamber capacities?

    The 1L capacity is best for nasal aerosols to maximize deposition below the top stage. The 2L capacity is optimized for nasal sprays to maximize nebulization and deposition. The 5L capacity is designed for stronger nasal sprays that require additional room to generate the aerosol plume.

    Is an expansion chamber required for Nasal Powder FPD testing?

    No, it is not mandatory. The test method for nasal powder is identical to that of dry powder inhalers, utilizing the same test instruments and accessories without requiring a pre-separator.

    What regulations and compliance standards do these testing methods follow?

    The testing procedures comply with FDA Bioavailability and Bioequivalence Studies (2003), FDA Nasal Spray, Inhalation Solution, Suspension & Spray Drug Products guidelines (2002), and Ph.Eur. 2021 (10.5) Nasal Preparations (Dosage Forms 0676).

    What parameters can the data analysis software calculate?

    The software automatically calculates the cutting particle size based on the flow rate. It also generates and displays key experimental parameters such as MMAD (Mass Median Aerodynamic Diameter) and GSD (Geometric Standard Deviation), alongside particle size distribution fitting curves.