Fully Automated Ames Test System HRH-AMES116 - Top China Suppliers & Factory for Bacterial Reverse Mutation Testing
Features
-
Experimental Throughput: 5 bacterial suspension channels and 14 sample channels. The sample channels can be arbitrarily configured for sample gradient concentrations, negative controls, or positive controls.
-
Solve the problems of hot hands and big errors exist in manual operation, achieve higher efficiency and more accurate dispensing.
-
Rapid Vortex Mixing: Use vortex mixing to rapidly homogenize the mixture.
-
Temperature Control for Reagent Storage: Realize top medium heating and S9 cooling with all temp adjustable.
-
Petri dish labeling system: Use inkjet coding to record and distinguish Petri dishes of different samples and different doses. After the experiment, a labeling reference table can be automatically generated.
-
Each parallel sample contains individual tubes and pipette tips. Waste pipette tips are isolated from the new one.
Application
Mutagenic tests for food, healthcare products, cosmetics, pharmaceuticals and new compounds.
Components
-
Metered Sampling and Pipetting
-
Rapid Vortex Mixing
-
Robot Arm Handling
-
Tube Distribution
-
Temperature Control
-
High-throughput Dish Basing
-
Bottom Medium Dispensing
-
Air Source
-
Host Control
HMI
Specification
| Petri Dish Throughput | 240 /test |
| Dish Replenish | Continuous and unlimited |
| Mixed Tube Throughput | 240 /test |
| Top Medium Dispensing Accuracy | ±0.3ml |
| Sample Adding Accuracy | ±0.01ml |
| Temp Control Range | 0 - 80℃ |
| Temp Control Accuracy | ±0.5℃ |
| Power Supply | 220V/50HZ |
| Host | All-in-one PC with industrial control |
| Number of Sample Gradient Concentration Processed | ≥5 /test |
| Number of Solvent Control Group Processed | ≥2 /test |
| Number of Positive Controls Placed | ≥6 /test |
| Number of Strains Treated | 5 /test |
| Compressed Air | oil-free at min. 5 bar |
| Communication Interface | Ethernet and special interface |
| Software | Dedicated control software with copyright |
| Dimensions | 1300mm×500mm×600mm |
| Net Weight | 55kg |
Standards
-
1. GB 15193.4-2014 National Food Safety Standard - Bacterial Reverse Mutation Test
-
2. YY/T 0127.10-2009 Biological evaluation of medical devices used in dentistry Part 2:Test method-Salmonella typhimurium reverse mutation assay (Ames mutagenicity test)
Frequently Asked Questions
What is the experimental throughput of this automated system?
The system features 5 bacterial suspension channels and 14 sample channels. The sample channels can be configured for sample gradient concentrations, negative controls, or positive controls. It delivers a throughput of 240 Petri dishes and 240 mixed tubes per test.
How does the system ensure temperature control accuracy?
The system realizes top medium heating and S9 cooling with a fully adjustable temperature range of 0 to 80℃, maintaining a precise temperature control accuracy of ±0.5℃.
What safety and hygiene measures are in place to prevent contamination?
Each parallel sample utilizes individual tubes and pipette tips. Furthermore, waste pipette tips are completely isolated from the new ones to ensure zero cross-contamination.
Which testing standards does this equipment comply with?
The system complies with GB 15193.4-2014 (National Food Safety Standard - Bacterial Reverse Mutation Test) and YY/T 0127.10-2009 (Biological evaluation of medical devices used in dentistry - Salmonella typhimurium reverse mutation assay / Ames mutagenicity test).
What are the primary applications of this mutagenic test device?
It is primarily used for conducting mutagenic tests for food, healthcare products, cosmetics, pharmaceuticals, and new compounds.
How does the system trace and distinguish different Petri dishes?
The system utilizes a Petri dish labeling system with inkjet coding to record and distinguish dishes of different samples and doses. After the experiment, a labeling reference table is automatically generated.