High-Flow Bioaerosol Sampler HRH-DSA350 from China Suppliers and Factory for Efficient Airborne Particle Collection
Features
Sampling Performance
- Sampling flow: 100-300L/min adjustable.
- Sampling efficiency: 70% when particle size ≤0.5μm.
- Sampling efficiency: 75% when particle size is 0.5-1μm.
- Sampling efficiency: 95% when particle size ≥2μm.
Operation & Control
- Sample form and volume: forms liquid sample onsite by eluent within 1 min, which can be used for subsequent analysis.
- Power duration: unintermittent operation 24h after fully charged.
- Control: supports wireless control and networking control.
- Sampling mode: four modes are available, including manual, auto, hardware triggered, and software triggered.
Physical Specifications
- Weight: sampler <3kg (with batteries).
- EMC: GJB151B-2013 compliant.
- Three-proof packaging: standard configurated three-proof packaging is easy for transportation.
Environmental Limits
- Working conditions: -40 ~ 70℃.
Software Functions
Software can set:
- a. sampling rate / b. sampling volume / c. sampling period and interval / d. sampling starts delay / e. number of sampling cycles / f. device number / g. network search.
- Real-time display: running time, number of completed cycles, sampled flow, remaining time, etc.
Application
Public & Custom Areas
- Public places such as bus stations, railway stations, subways, airports, schools, etc.
- Customs and frontier fields.
Health, Food & Agriculture
- Public health areas such as indoor air quality in clean rooms, hospitals, clean workshops, etc.
- Food processing fields.
- Livestock farms, breeding fields, farmers' markets, etc.
Working Conditions
- Temperature: -40℃ ~ 70℃
- Conditions: all-weather (optional rain cover)
Storage Environment
- Temperature: -40℃ ~ 70℃
- Humidity: 5% ~ 98% RH (non-condensing state)
Sampler Technical Parameters
| Sampling range | Suitable for collecting biological aerosols, particulate matter, submicron particles, such as bacteria, viruses, fungi, molds, pollen and phage, etc. The standard bioaerosol cartridge is used for collecting bioaerosols and particulate aerosols. |
| Membrane type | Standard configurated biological aerosol cartridges. |
| Sampling mode | Four modes are available, including manual, auto, hardware triggered, software triggered. |
| Sampling principle | Filter trapping |
| Sampling flow | 100-300L/min adjustable |
| Sampling efficiency |
Sampling efficiency ≥70% when particle size ≤0.5μm. Sampling efficiency ≥75% when particle size is 0.5-1μm. Sampling efficiency ≥95% when particle size ≥2μm. |
| Sample form and volume | Form liquid sample onsite by eluent within 1 min, which can be used for subsequent analysis. |
| Working temp | -40℃ ~ 70℃ (work uninterruptedly for 2hrs under extreme temp) |
| Storage temp | -40℃ ~ 70℃ |
| Relative humidity | 98% (35 ℃) |
| Working conditions | All-weather conditions (special protection kit for rain and snow) |
| Power supply | Built-in rechargeable battery, 110V/220V/50-60Hz AC or external 12V or 24V DC. Power rate <10W. |
| Power duration | Unintermittent operation ≥ 30h after fully charged (sampling efficiency ≥300L/min). |
| Sampling pump (fan) service life | 60,000 hrs. |
| Noise | ≤ 60dB |
| Weight | Sampler <3kg (with batteries) |
| Casing | The main body of the casing is made of aluminum alloy, which is anti-electromagnetic and waterproof. |
| Control mode | Wired and wireless control |
| Interface | Standard DB-9 port |
| Software | Each sampler is equipped with software, software can set a. sampling rate b. sampling volume c. sampling period and interval d. sampling starts delay e. number of sampling cycles f. device number g. network search. Real-time display: running time, number of completed cycles, sampled flow, remaining time, etc. |
| Bioaerosol alarm trigger function | Can be set and worked together with bioaerosol alarms. When bioaerosol alarms, the high-flow bioaerosol sampler will be activated, the sampler will start work. |
| Sampler networking | Samplers can be connected to network, multiple samplers in the network can be monitored and controlled at any time. |
| Cleaning and Sterilization | Fan housing and bioaerosol filter can be sterilized after wholly disassembling, and the sampler can be partially immersed in disinfectant or wiped on the surface. |
Eluent Technical Parameters
| Principle | High-frequency vibration and mechanical pressure of membrane in liquid. |
| Elution time | ≤1min |
| Power supply | 4x AA batteries or 220VAC 50Hz. |
| Weight | 1kg |
Frequently Asked Questions
What is the sampling flow rate and is it adjustable?
The sampling flow rate is fully adjustable from 100 to 300 L/min, allowing users to customize the flow based on specific environmental testing needs.
What is the sampling efficiency for different particle sizes?
The system features high collection efficiency: ≥70% for particles ≤0.5μm, ≥75% for particles between 0.5-1μm, and ≥95% for larger particles ≥2μm.
How long does the battery last on a single charge?
The built-in rechargeable battery supports unintermittent operation for ≥30 hours after being fully charged, even when running at a high sampling flow rate of ≥300L/min.
Can the sampler be controlled remotely?
Yes, the sampler supports both wired and wireless control. It also features networking capabilities, allowing multiple devices to be monitored and controlled simultaneously.
What are the working and storage temperature limits?
Both the working temperature and storage temperature range from -40℃ to 70℃. The system can work uninterruptedly for 2 hours under extreme temperatures.
How does the eluent device work and how fast is it?
The eluent system operates on high-frequency vibration and mechanical pressure of the membrane in liquid. It forms a liquid sample onsite within 1 minute, which is ready for subsequent analysis.