HRH-WS300 Environmental Simulation Chamber - Reliable Temperature & Humidity Control by China Suppliers and Factory
Application
Orally Inhaled & Nasal Drug Products
Features
Specially designed and customized for the testing of oral inhalers and nasal drug products.
Temp and humidity control is accurate, repeatable and reproducible, and the temp and humidity inside the chamber are uniformly distributed.
With static electricity elimination system, minimizing the impact of static charges on the experiment. The static electricity elimination function turns on/off automatically along with the actual temperature and humidity inside the chamber.
Easy and convenient operation with touch screen control.
The lighting system is convenient for operation.
No daily maintenance required, increasing users' satisfaction.
Can be used together with the cascade impactor cooler to minimize the atmosphere impact on the inhaler testing.
USP and Ph. Eur. compliant.
Specification
| Temperature control range | 17-35℃ |
| Temperature uniformity | ±2℃ (no load) |
| Humidity control range | 15%-90% RH |
| Humidity uniformity | ±5% RH (no load) |
| Heating rate (mean value) | 1-4℃ /min (no load) |
| Cooling rate (mean value) | 0.7-1.2℃ /min (no load) |
| Standard Configuration | Electrostatic minimization system |
| Screen control | Touch panel |
| Max. dimensions of chamber (W x D x H) | 1000mm x 490mm x 590mm |
| Overall dimensions (W x D x H) | 1400mm x 810mm x 1060mm |
| Power rate | 5.5kw |
| Power supply | 220V 50Hz 16A |
| Interfaces | Left side: cable, pneumatic; Right side: equipment, cable, pneumatic |
Frequently Asked Questions
Q: What are the primary applications for this testing chamber?
This chamber is specially designed and customized for the testing of Orally Inhaled & Nasal Drug Products (OINDPs).
Q: What are the temperature and humidity control ranges of the system?
The temperature control range is 17-35℃ (with a uniformity of ±2℃ under no load conditions), and the humidity control range is 15%-90% RH (with a uniformity of ±5% RH under no load conditions).
Q: How does the chamber handle static electricity during testing?
It features a built-in electrostatic minimization system that automatically turns on and off along with the actual temperature and humidity inside the chamber, minimizing static charge impacts on your experiments.
Q: Does this testing chamber comply with international pharmacopoeia standards?
Yes, the testing chamber is fully compliant with both USP (United States Pharmacopeia) and Ph. Eur. (European Pharmacopoeia) standards.
Q: Can it be integrated with other testing accessories like cascade impactor coolers?
Yes, it can be used in combination with a cascade impactor cooler to minimize the atmospheric impact on the inhaler testing process.
Q: What are the power supply requirements and physical dimensions?
The equipment requires a 220V 50Hz 16A power supply with a power rate of 5.5kw. The overall dimensions are 1400mm x 810mm x 1060mm, and the maximum chamber dimensions are 1000mm x 490mm x 590mm.