Nose-Only Exposure Research Solutions

2026-06-08

Nose-only exposure systems are the primary method for acute inhalation exposure. Two flow designs exist: Directed Flow and Past Flow. OECD Test Guidelines 403 and 436 recommend Directed Flow as the preferred method because it minimizes dilution of exposure concentration by exhaled air and reduces interference between animals. Most acute inhalation toxicity studies now follow the Directed Flow design. OECD TG 403 and 436 specify exposure parameters: temperature 22±3°C, humidity 30–70%, CO₂ ≤1%, O₂ ≥19%. HRHINS rodent nose-only exposure system (Directed Flow) is widely used in safety evaluations of pesticides, chemicals, and new drugs. It monitors these parameters in real time, integrates with dust and liquid aerosol generators, and ensures that MMAD is maintained at 1–4 µm, GSD at 1.5–3, with aerosol concentration stability within ±20% throughout exposure.