Pathogenic Bioaerosol Infection System for Virus Research - China Suppliers & Factory Solutions
Features
- It can perform bioaerosol infection tests on rodents and medium-sized animals e.g., monkeys, guinea pigs, ferrets, rats, and mice.
- Three infection tests of direct contact transmission, droplet transmission and aerosol transmission can be performed.
- Meet long-term feeding requirements of laboratory animals.
- With rich functions of environmental monitoring and biosampling.
- With smart control software and system.
- For biosafety protective tightness requirements of Class III BSC cabinet working area, it conforms to RB/T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment.
- Class III BSC cabinet adopts HEPA filtration for air inlet and double filtration for outlet. HEPA filter has a sterilization port and leakage detection port.
- The corner of Class III BSC cabinet adopts large arc design with sterilization port.
Application
Studies of pathogenic bioaerosol infections
Specification
| Applicable animals | Monkeys, guinea pigs, ferrets, rats, mice (as requested) |
| Droplet transmission distance | 5cm |
| Aerosol transmission distance | 1.5m |
| Number of animals in direct contact transmission group | 4-5 guinea pigs (other animals can be customized) |
| Number of animals in droplet transmission group | 1-2 guinea pigs (other animals can be customized) |
| Number of animals in aerosol transmission group | 1-2 guinea pigs (other animals can be customized) |
| Transmitted airflow | 20-100L/min adjustable |
| Bio-sampling flow | 28.3L/min |
| Environmental parameters monitoring | Oxygen, carbon dioxide, ammonia, temp, humidity, etc. |
Standards
RB∕T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment
GB 14925-2010 Laboratory animal-Requirements of environment and housing facilities
Frequently Asked Questions
What types of animals can be tested with this system?
The system is designed to perform bioaerosol infection tests on rodents and medium-sized animals. This includes monkeys, guinea pigs, ferrets, rats, and mice as requested.
Which infection transmission modes can be tested?
The equipment supports three distinct infection transmission tests: direct contact transmission, droplet transmission, and aerosol transmission.
How does the Class III BSC cabinet ensure biosafety tightness?
It complies with the RB/T 199-2015 biosafety performance standards. It features HEPA filtration for the air inlet and double filtration for the outlet (equipped with sterilization and leakage detection ports), and the cabinet corners utilize a large arc design with a sterilization port.
What are the default transmission distances for droplet and aerosol tests?
The droplet transmission distance is set at 5cm, while the aerosol transmission distance is set at 1.5m.
What environmental parameters can the system monitor?
The system provides rich environmental monitoring capabilities, tracking parameters such as oxygen, carbon dioxide, ammonia, temperature, and humidity.
What standards does this equipment comply with?
The equipment complies with the RB∕T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment and GB 14925-2010 Laboratory animal-Requirements of environment and housing facilities.