Pathogenic Bioaerosol Infection System - Leading China Suppliers & Factory for Virus Research in Rodents & Medium-Sized Animals
Features
- It can perform bioaerosol infection tests on rodents and medium-sized animals e.g., monkeys, guinea pigs, ferrets, rats, and mice.
- Three infection tests of direct contact transmission, droplet transmission and aerosol transmission can be performed.
- Meet long-term feeding requirements of laboratory animals.
- With rich functions of environmental monitoring and biosampling.
- With smart control software and system.
- For biosafety protective tightness requirements of Class III BSC cabinet working area, it conforms to RB/T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment.
- Class III BSC cabinet adopts HEPA filtration for air inlet and double filtration for outlet. HEPA filter has a sterilization port and leakage detection port.
- The corner of Class III BSC cabinet adopts large arc design with sterilization port.
Application
- Studies of pathogenic bioaerosol infections
Specification
| Applicable animals | Monkeys, guinea pigs, ferrets, rats, mice (as requested) |
| Droplet transmission distance | 5cm |
| Aerosol transmission distance | 1.5m |
| Number of animals in direct contact transmission group | 4-5 guinea pigs (other animals can be customized) |
| Number of animals in droplet transmission group | 1-2 guinea pigs (other animals can be customized) |
| Number of animals in aerosol transmission group | 1-2 guinea pigs (other animals can be customized) |
| Transmitted airflow | 20-100L/min adjustable |
| Bio-sampling flow | 28.3L/min |
| Environmental parameters monitoring | Oxygen, carbon dioxide, ammonia, temp, humidity, etc. |
Standards
- RB∕T 199-2015 Technical code for evaluating biosafety performance of laboratory equipment
- GB 14925-2010 Laboratory animal-Requirements of environment and housing facilities
Frequently Asked Questions
Which animal species are compatible with this bioaerosol infection testing system?
The system is designed to perform bioaerosol infection tests on rodents and medium-sized animals, including monkeys, guinea pigs, ferrets, rats, and mice, with customization options available for other species.
What types of transmission pathways can be simulated and tested?
It supports three distinct infection transmission testing modes: direct contact transmission, droplet transmission (at a 5cm distance), and aerosol transmission (at a 1.5m distance).
How does the system ensure biosafety and containment?
The cabinet meets Class III BSC tightness requirements in accordance with RB/T 199-2015. It features HEPA filtration for the air inlet, double filtration for the outlet, and includes dedicated sterilization and leakage detection ports.
What environmental parameters are monitored during testing?
The system provides comprehensive monitoring of critical environmental parameters inside the cabinet, including oxygen, carbon dioxide, ammonia levels, temperature, and relative humidity.
What biosafety standards does this equipment comply with?
It strictly conforms to RB/T 199-2015 (Technical code for evaluating biosafety performance of laboratory equipment) and GB 14925-2010 (Laboratory animal-Requirements of environment and housing facilities).