Request a Quote
Leave Your Message

Pulmonary Function Testing System for Rodents by Professional China Suppliers and Manufacturers - Inhalation Toxicity Testing Solution

The Pulmonary Function Testing System for Rodents is an advanced solution designed for comprehensive inhalation toxicity testing. This state-of-the-art system operates seamlessly with the nose-only exposure system to accurately measure, analyze, display, and record critical pulmonary function parameters in rodents. As a leading factory in China, we pride ourselves on providing high-quality equipment that meets the needs of researchers and laboratories worldwide. Our reliable products and commitment to excellence make us a preferred choice among suppliers in the industry. Enhance your research capabilities with our innovative Pulmonary Function Testing System for Rodents today

    Features

    • Real-time monitoring of rodent breathing parameters while performing exposure tests.
    • Easy to operate, no surgery required.
    • Same rodent comparison before and after drug administration.
    • Suitable for long-term tracking research.
    • Applicable to mice, rats, guinea pigs, etc.

    Application

    Rodent Physiology Studies, Rodent Screening, Drug Safety Assessment.

    Major Components

    1 Control cabinet

    The control cabinet mainly consists of a data acquisition card and a linear vacuum pump, which is used to collect plethysmograph signals detected by sensors and provide negative pressure airflow for ventilation at a certain flow rate. The detected signals are transmitted to data analysis software via a USB cable connected to the computer.

    2 Animal plethysmograph

    The animal plethysmograph is used to hold animals to be tested, which shall be kept dry inside before use and cleaned up in time after use.

    3 Pressure sensor

    The pressure sensor is used for detecting pressure changes caused by animal breathing in plethysmographs. It is connected to an animal plethysmograph and control cabinet.

    4 Desiccant

    Desiccant is connected between ventilation ports among the plethysmograph and the control cabinet, which is used for drying gas extracted from the plethysmograph. The desiccant in the filter adopts silica gel with a moisture indicator, which needs to be replaced when it turns pink.

    5 Data analysis system

    The data analysis system is a multifunctional system of data acquisition, analysis, display and storage, specially designed for the WBP system.

    Specification - Parameters

    Parameters Meaning
    Ti Inspiratory time (ms)
    Te Expiratory time (ms)
    f Respiratory rate (bpm)
    TV Tidal volume (mL)
    MV Minute ventilation (mL)
    RT Relaxation time (ms)
    EV Expiratory volume (mL)
    PIF Peak inspiratory flow (mL/min)
    PEF Peak expiratory flow (mL/min)
    EEP End-expiratory pause (ms)
    Tbox Body box temperature (°C)
    Tbody Animal body temperature (°C)
    Patm Atmospheric pressure (mmHg)
    VCF Correction factor
    Penh Enhanced pause / bronchoconstriction index
    EF50 50% expiratory flow rate (mL/min)
    EIP End-inspiratory pause (ms)

    Data Analysis Performance

    • 1. Real-time collection and analysis curve parameters, display the average value of respiratory parameters within 5s.
    • 2. Historical test data query, playback at the fastest 32x speed.
    • 3. Analysis data can be exported to Excel.
    Data Analysis Performance1

    Specification - Hardware

    Item Channel 4 Channel 8
    Communication port USB type-D USB type-D
    Sensor range ±2.5 H2O ±2.5 H2O
    Power supply 220VAC 50Hz 220VAC 50Hz
    Weight 13.3kg 15.5kg
    Dimensions 618x320x230mm 618x320x230mm

    Working & Storage Conditions

    Working Conditions

    Temp: 0℃~60℃
    Humidity: 5%~95% (non-condensing)

    Storage Conditions

    Temp: -20℃~70℃
    Humidity: 5%~95% (non-condensing)

    Conditions (Publications)

    • [1] Wang Guanlin, Zhai Zizhao, Wang Wenhao, et al. Tailored Borneol-Modified Lipid Nanoparticles Nasal Spray for Enhanced Nose-to-Brain Delivery to Central Nervous System Diseases. ACS Nano, 2024, https://doi.org/10.1021/acsnano.4c08279. (IF 15.8)
    • [2] Linjing Wu, Wenhao Wang, Mengqin Guo, et al. Inhalable iron redox cycling powered nanoreactor for amplified ferroptosis-apoptosis synergetic therapy of lung cancer. Nano Res. 2024, https://doi.org/10.1007/s12274-024-6455-2. (IF 9.5)

    Frequently Asked Questions

    Q: Which animal subjects are suitable for this plethysmograph system?

    The system is designed for rodent research and is highly applicable to mice, rats, guinea pigs, and similar species.

    Q: Does the system require surgical operation on the animals?

    No, the system is completely non-invasive and easy to operate. There is no need for surgical procedures on the rodents.

    Q: What parameters can be tracked in real-time?

    It monitors various physiological parameters in real-time, including Inspiratory/Expiratory time (Ti/Te), Respiratory rate (f), Tidal volume (TV), Minute ventilation (MV), and the Enhanced pause (Penh) bronchoconstriction index, among others.

    Q: How do I know when the desiccant needs to be replaced?

    The desiccant in the filter uses silica gel with a moisture indicator. You should replace it immediately when the silica gel turns from its original color to pink.

    Q: Can I export the collected data for external analysis?

    Yes, the integrated data analysis system allows you to export the analyzed data directly to Excel formats, as well as query and play back historical test data at up to 32x speed.

    Q: What configurations are available for the system channels?

    The system is available in 4-channel and 8-channel configurations to meet different laboratory throughput requirements.

    Make An AppointmentLEAVE YOUR MESSAGE