Single-Channel Intelligent Smoke Generator from China Suppliers and Factory - Model HRH-SM120 Meeting ISO Standards
Features
- Load with two cigarettes simultaneously.
- Smoke capacity: 15-60ml.
- Smoke cycle: 10-240s.
- Smoke curve: standard bell-shaped curve or other customized curves.
- Measurement capacity accuracy: ±0.01 mL
- Control capacity accuracy: ±0.01 mL
- Measurement time accuracy: ±0.01 s
- Control time accuracy: ±0.01 s
- Puff duration: 3–60 s adjustable
- Rated voltage: 220 V
- Main unit weight: 5 kg
- Main unit dimensions: 37 × 23 × 38 (cm)
Working Environment
- Temperature: 0 ℃–60 ℃
- Humidity: 5%–95% (non-condensing)
Storage Environment
- Temperature: -20 ℃–70 ℃
- Humidity: 5%–95% (non-condensing)
Routine Maintenance
- If the tar cannot be cleaned thoroughly after prolonged experiments, remove the syringe, pull out the plunger, and clean it with a soft test tube brush. After cleaning, apply an appropriate amount of Vaseline around the plunger base (the white front-end part) for lubrication.
- The tubing used in this product is 4x6mm peristaltic pump tubing, with a hardness of approximately Shore 55A. During experiments, it shall be replaced regularly. Compliant tubing can be procured independently or purchased through our Company's sales department.
- The maximum service life of the pinch valve used in this product is 10 million cycles. Replace it in advance to avoid affecting the experiment.
Pinch Valve Replacement Steps
1
Remove the silicone tubing from the pinch valve.
2
Use a hex wrench to remove the four screws.
3
Disconnect the wiring of the pinch valve (a plug-in terminal connector is used here; press the middle protrusion and pull firmly apart).
4
Finally, use a hex wrench to remove the screws on the pinch valve, and then remove the pinch valve.
5
Install the new pinch valve by following the reverse order of Steps 4-3-2-1.
Application
Applicable to various toxicology research centers, animal smoking poisoning tests, respiratory laboratories of hospitals, technical laboratories of cigarette factories for smoke research, etc.
Frequently Asked Questions (FAQ)
Q1: What is the cigarette capacity of this device?
The device can be loaded with two cigarettes simultaneously, featuring a adjustable smoke capacity of 15 to 60 ml and a smoke cycle range of 10 to 240 seconds.
Q2: What are the working and storage environment requirements?
The working environment temperature should be between 0 ℃ and 60 ℃. For storage, the temperature range is -20 ℃ to 70 ℃. In both states, the humidity must remain between 5% and 95% (non-condensing).
Q3: How do I clean and maintain the syringe and plunger?
If tar cannot be cleaned thoroughly, remove the syringe, pull out the plunger, and clean it with a soft test tube brush. After cleaning, apply an appropriate amount of Vaseline around the white front-end part of the plunger base for lubrication.
Q4: What type of tubing is required, and how often should it be replaced?
The device utilizes 4x6mm peristaltic pump tubing with a hardness of approximately Shore 55A. It needs to be replaced regularly during experiments and can be procured independently or via our sales department.
Q5: What is the lifespan of the pinch valve, and how do I replace it?
The pinch valve has a maximum service life of 10 million cycles. To replace it, remove the silicone tubing, unscrew the four screws with a hex wrench, disconnect the wiring plug-in terminal, remove the valve, and install the new one in reverse order.
Q6: What fields and industries is this equipment designed for?
This equipment is widely applicable to toxicology research centers, animal smoking poisoning tests, hospital respiratory laboratories, and technical laboratories of cigarette factories for smoke research.