Two-Fluid Liquid Aerosol Generator - High Precision Gas-Liquid Control from Leading China Suppliers and Factory
Features
Application
It can be applied to inhalation toxicology research, aerobiology, aerosol research, biohazard testing, performance evaluation of filters, inhalation immunization, inhalation therapy, drug research, environmental evaluation, hazard assessment of biology and chemistry, and medical protection. Especially suitable for viscous or heavy oils which are difficult to be generated by traditional generators, also can be used for various water-soluble, oil-soluble and part of the fine particles stabilized suspension and other solutions samples.
Aerosol Generation
After connecting the compressed air pipeline and aerosol pipeline, open the compressed air valve, the compressed air inside the two-flow nozzle forms high-speed rotary flow, resulting in a strong vortex circumstance, the liquid sample is pumped in by the peristaltic pump and then dispersed mixed with high-speed spray to form aerosols. Big particles impact to the buffer chamber wall or deposit in the forward process, and discharged through the waste liquid outlet, small particles will be output from the aerosol outlet.
Specification
| Parameter | Specification |
|---|---|
| Specification of silicon tube in connection with liquid tube | 2mm x 4mm |
| Outer diameter of PU tubes in connection with air compressor | 4mm x 6mm |
| Aerosol outlet specification | 16mm x 20mm |
Standards
Frequently Asked Questions (FAQ)
What makes the HRH-TAG1 aerosol generator suitable for viscous or heavy oils?
Unlike traditional aerosol generators, the HRH-TAG1 features an advanced two-flow nozzle design that creates a strong high-speed rotary vortex flow. This enables the effective dispersion and atomization of viscous or heavy oils that are typically difficult to generate.
Can the HRH-TAG1 support long-term continuous inhalation experiments?
Yes. The system can be connected to an external peristaltic pump to continuously add liquid samples, meeting the demands of long-term and stable experimental aerosol generation.
What is the particle size distribution generated by this device?
The HRH-TAG1 generates aerosols with a Mass Median Aerodynamic Diameter (MMAD) of 1-3μm and a Geometric Standard Deviation (GSD) of 1.5-3, making it ideal for targeted respiratory and inhalation research.
Is the particle size buffer chamber easy to maintain?
Yes, it features a streamlined, removable design for the particle size sieving buffer chamber, allowing for quick disassembly, easy cleaning, and efficient maintenance.
Which international testing standards does the HRH-TAG1 comply with?
It is fully compliant with key international inhalation toxicology guidelines, including OECD TG 403 (Acute Inhalation), OECD TG 436 (Acute Toxic Class Method), OECD TG 412 (28d Sub-acute Inhalation), and OECD TG 413 (90d Sub-chronic Inhalation).
Can the HRH-TAG1 be integrated with animal exposure systems?
Yes, it is designed for easy integration and can be seamlessly connected with our nose-only exposure systems for rodent studies.