VOC Generator
Features
Gas concentration generated is controllable and with high continuous stability.
Adopt glass ball stack convective volatilization structure with fast volatilization speed.
Assist heating volatile gas, temp can be controlled.
Generating system is closed with high safety.
Volatilization rate: 0-5mL/min, gas flow rate: 0-20L/min, heating temp: 20-60℃.
Application
Inhalation Toxicity Tests on VOC.
Health Risk Assessment of Environmental Atmosphere and Indoor Volatile Solvents.
Inhalation Toxicity Studies of VOC or Petrochemicals in Occupational Sites.
Online Monitor and Control for VOC.
Specification
| Volatilization rate | 0-5mL/min |
| Gas flow rate | 0-20L/min |
| Heating temperature | 20-60℃ |
Working principle
The volatile sample is quantitatively injected into the volatilization pipeline filled with glass beads through the sample transfer pump, the superficial area of the dropped sample is increased enough to allow for faster volatilization and the clean air is heated by the air heater to enter the cushion chamber, and then the sample volatilization is promoted by the glass sphere pile and volatilized samples and air can be imported into the exposure system through the outlet for contamination. There is a temperature probe at the outlet of VOC gas for detecting the temperature of the outlet gas, the temperature value is fed back to the air heater, and the heater is used to intelligently adjust the heating power according to the outlet temperature set by the user.

Note: 1. VOC concentration = liquid sample mass/ (liquid sample gasification expansion volume + air volume);
When the experimental concentration of VOC is designed to be relatively low, the concentration of VOC is ≈ liquid sample mass/air volume;
2. The temperature value set by the air heater is the temperature of the air outlet of the VOC system other than the hot air temperature after heating;
3. To be able to quantify the VOC content, the injection rate of the sample cannot exceed the maximum volatilization rate, that's to say that it's necessary to ensure that the sample injected into the volatilization tube is completely volatilized, and no liquid flows into the cushion chamber.
4. The temperature setting method is as follows: The temperature control mode is that the system will automatically stop the air heating when the air outlet temperature is greater than the set upper limit; the system will automatically enable the air heating when the temperature is less than the set lower limit.
When the experimental concentration of VOC is designed to be relatively low, the concentration of VOC is ≈ liquid sample mass/air volume;
2. The temperature value set by the air heater is the temperature of the air outlet of the VOC system other than the hot air temperature after heating;
3. To be able to quantify the VOC content, the injection rate of the sample cannot exceed the maximum volatilization rate, that's to say that it's necessary to ensure that the sample injected into the volatilization tube is completely volatilized, and no liquid flows into the cushion chamber.
4. The temperature setting method is as follows: The temperature control mode is that the system will automatically stop the air heating when the air outlet temperature is greater than the set upper limit; the system will automatically enable the air heating when the temperature is less than the set lower limit.